MRU600

Portable, stand alone Gas Conditioning System High quality flue gas conditioning guarantees highly accurate measuring results. Condensation of the flue gas sample must be avoided in the pathway from the point of gas extraction to the flue gas analyzer. None of the target substances must be lost due to condensation. The gas conditioning system protects the analyzer against humidity and dust particles. Thus, long term measuring over a period of several hours becomes possible.

    • 100% corrosion resistance of the gas pathways
    • Option: No condensation on the way from the gas extraction point due to heated sampling line
    • Temperatur regulation to protect the physical condition of the flue gas sample
    • More than 99% efficiency guaranteed
    • Condensation of the flue gas humidity in the heat exchanger at 5° C
    • Indication of the outlet dew point temperature on the front panel of the unit
    • Individually adjustable dew point temperature

Flow
50 ... 150l/h or 50 ... 250l/h (w/ double heat exchanger)
Ambient air temperature
+5°C ... +50°C

Flow
50 ... 150l/h or 50 ... 250l/h (w/ double heat exchanger)
Ambient air temperature
+5°C ... +50°C

Cooling capacity
100 kj/h
Inlet dew point
40°C up to 65°C max. (depending on flow rate)
Outlet dew point
+ 5°C, Alarm at deviation of ± 3°C
Cooler type
Peltier controlled heat exchanger
Power consumption
max. 120W
Warm up time
appr. 5 minutes
Dimensions
410x360x210 mm (WxHxD)
Weight
10 kg